Potentiometric Sensor for Benzylpenicillin Determination


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A potentiometric benzylpenicillin-sensitive sensor with a plasticized polyvinyl-chloride membrane was developed. The sensor contains a benzylpenicillin—rhodamine 6G ion pair. The response is linear for benzylpenicillin concentrations in the range 10–3 – 10–1M with an electrode-function slope characteristic of singly charged anions and a minimum detected concentration of (6.3 – 7.9) × 10–4 M. This sensor could be used to detect and determine benzylpenicillin in pharmaceuticals.

作者简介

Zh. Kormosh

Lesya Ukrainka Eastern European National University

编辑信件的主要联系方式.
Email: zholt-1971@ukr.net
乌克兰, Lutsk, 43025

N. Markovska

Lesya Ukrainka Eastern European National University

Email: zholt-1971@ukr.net
乌克兰, Lutsk, 43025

N. Kormosh

Lutsk Basic Medical College

Email: zholt-1971@ukr.net
乌克兰, Lutsk, 43016

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media, LLC, part of Springer Nature, 2019