Methods for Compensation of Coil Misalignment in Systems for Inductive Transcutaneous Power Transfer to Implanted Medical Devices


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

The problem of variation of the output characteristics (voltage, current, power) of inductive powering systems for implantable medical devices is considered. The variation is primarily caused by relative misalignment of transmitting and receiving coils. A classification of misalignments is proposed. Main approaches to coil misalignment compensation are: mechanical fixation of the coils; evaluation of the relative position of the coils for elimination of misalignment or further tuning of the transmitting part of the system; optimization of the coil geometry providing stability within the given misalignment range; tuning the working frequency of the transmitting part of the system. It is shown that the compensation method should be selected with due regard to the implantable device parameters (first of all, the coil implantation site and the power consumption of the device).

Sobre autores

A. Danilov

National Research University of Electronic Technology

Autor responsável pela correspondência
Email: arseny.danilov@gmail.com
Rússia, Zelenograd, Moscow

E. Mindubaev

National Research University of Electronic Technology

Email: arseny.danilov@gmail.com
Rússia, Zelenograd, Moscow

S. Selishchev

National Research University of Electronic Technology

Email: arseny.danilov@gmail.com
Rússia, Zelenograd, Moscow

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer Science+Business Media New York, 2017