Methods for Compensation of Coil Misalignment in Systems for Inductive Transcutaneous Power Transfer to Implanted Medical Devices
- 作者: Danilov A.A.1, Mindubaev E.A.1, Selishchev S.V.1
-
隶属关系:
- National Research University of Electronic Technology
- 期: 卷 51, 编号 1 (2017)
- 页面: 56-60
- 栏目: Article
- URL: https://ogarev-online.ru/0006-3398/article/view/234897
- DOI: https://doi.org/10.1007/s10527-017-9684-9
- ID: 234897
如何引用文章
详细
The problem of variation of the output characteristics (voltage, current, power) of inductive powering systems for implantable medical devices is considered. The variation is primarily caused by relative misalignment of transmitting and receiving coils. A classification of misalignments is proposed. Main approaches to coil misalignment compensation are: mechanical fixation of the coils; evaluation of the relative position of the coils for elimination of misalignment or further tuning of the transmitting part of the system; optimization of the coil geometry providing stability within the given misalignment range; tuning the working frequency of the transmitting part of the system. It is shown that the compensation method should be selected with due regard to the implantable device parameters (first of all, the coil implantation site and the power consumption of the device).
作者简介
A. Danilov
National Research University of Electronic Technology
编辑信件的主要联系方式.
Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow
E. Mindubaev
National Research University of Electronic Technology
Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow
S. Selishchev
National Research University of Electronic Technology
Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow
补充文件