Methods for Compensation of Coil Misalignment in Systems for Inductive Transcutaneous Power Transfer to Implanted Medical Devices


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

The problem of variation of the output characteristics (voltage, current, power) of inductive powering systems for implantable medical devices is considered. The variation is primarily caused by relative misalignment of transmitting and receiving coils. A classification of misalignments is proposed. Main approaches to coil misalignment compensation are: mechanical fixation of the coils; evaluation of the relative position of the coils for elimination of misalignment or further tuning of the transmitting part of the system; optimization of the coil geometry providing stability within the given misalignment range; tuning the working frequency of the transmitting part of the system. It is shown that the compensation method should be selected with due regard to the implantable device parameters (first of all, the coil implantation site and the power consumption of the device).

作者简介

A. Danilov

National Research University of Electronic Technology

编辑信件的主要联系方式.
Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow

E. Mindubaev

National Research University of Electronic Technology

Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow

S. Selishchev

National Research University of Electronic Technology

Email: arseny.danilov@gmail.com
俄罗斯联邦, Zelenograd, Moscow

补充文件

附件文件
动作
1. JATS XML

版权所有 © Springer Science+Business Media New York, 2017