Effect of Eluting the 99Mo/99mTc Generator More than Once in 24 h on the Efficiency of the 99mTc-MIBI Complex
- 作者: Hassanpour S.H.1, Dehghani M.A.2, Elhamasl R.3, Lalehzari R.4
-
隶属关系:
- Yasooj Branch
- Department of Toxicology, Faculty of Pharmacy
- Department of Medical Physics, Faculty of Medicine
- Faculty of Civil Engineering
- 期: 卷 61, 编号 1 (2019)
- 页面: 73-80
- 栏目: Article
- URL: https://ogarev-online.ru/1066-3622/article/view/224921
- DOI: https://doi.org/10.1134/S1066362219010119
- ID: 224921
如何引用文章
详细
The effect of heating time, heating temperature, the amount of activity added to the MIBI kit, and the age of the formulated kit on the radiochemical purity of 99mTc-MIBI complex is demonstrated. 1-mCi (37-MBq) 99mTc-MIBI samples were prepared using the standard boiling water bath procedure. Qualitative and quantitative studies were done for the samples. The radiochemical purity of 99mTc that had been eluted from two types of 99Mo/99mTc generators was studied. The efficiency of 99mTc-MIBI samples was tested based on these studies. The results obtained show that eluting 99Mo/99mTc generators more than once in 24 h can lead to increased medical risks in patients with myocardial infarction.
作者简介
S. Hassanpour
Yasooj Branch
编辑信件的主要联系方式.
Email: Dr.hossein1366@yahoo.com
伊朗伊斯兰共和国, Yasooj
M. Dehghani
Department of Toxicology, Faculty of Pharmacy
Email: Dr.hossein1366@yahoo.com
伊朗伊斯兰共和国, Ahvaz
R. Elhamasl
Department of Medical Physics, Faculty of Medicine
Email: Dr.hossein1366@yahoo.com
伊朗伊斯兰共和国, Shiraz
R. Lalehzari
Faculty of Civil Engineering
Email: Dr.hossein1366@yahoo.com
伊朗伊斯兰共和国, Tehran
补充文件