Determination of Cycloserine in Blood Plasma by HPLC/MS: Application to Bioequivalence Studies


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

A selective and specific quantitative assay with HPLC/MS determination of the broad-spectrum antituberculosis antibiotic cycloserine in blood plasma with reserpine internal standard was developed and validated. It was shown that the method was accurate and precise and linear in the range 0.5 – 20.0 μg/mL. A calibration curve obeyed the equation Y = aX + b, where a = 0.0893 and b = –0.0134 with a correlation coefficient of 0.9994. The cycloserine limit of detection was 0.1 μg/mL; the limit of quantitation, 0.5 μg/mL. The recoveries of cycloserine from the matrix at low (0.5 μg/mL) and high (20.0 μg/mL) drug concentrations were 41.7 and 62.4%, respectively, with matrix factors of 0.28 and 0.34, respectively, normalized to the internal standard.

Авторлар туралы

E. Stepanova

Shared Research and Educational Center, People’s Friendship University of Russia

Email: chem@folium.ru
Ресей, Moscow, 117198

M. Ovcharov

Shared Research and Educational Center, People’s Friendship University of Russia

Email: chem@folium.ru
Ресей, Moscow, 117198

S. Barsegyan

Shared Research and Educational Center, People’s Friendship University of Russia

Email: chem@folium.ru
Ресей, Moscow, 117198

V. Chistyakov

Shared Research and Educational Center, People’s Friendship University of Russia

Email: chem@folium.ru
Ресей, Moscow, 117198

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Springer Science+Business Media New York, 2016