Certification of Reference Standards in Pharmacy: Mass Balance Method


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

Approaches to certification of substances as reference standards (RSs) by the mass-balance method and the restrictions that must be considered when using this method are considered. Formulas for estimating the uncertainty of certified values are presented. Also, formulas missing from the literature are given to assess a) the uncertainty in the contents of organic impurities obtained by normalization of peak areas (heights) with correction factors and b) the uncertainty in the detection limits of the organic impurities. Recommendations for correct certification of reference standards by the mass-balance method are provided.

About the authors

N. A. Epshtein

STADA PharmDevelopment R&D Corporation, NIZHPHARM Company

Email: chem@folium.ru
Russian Federation, 6 Avtomobil’nyi Proezd, Moscow, 109029

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2018 Springer Science+Business Media, LLC, part of Springer Nature