Biomedical Optoacoustic Tomograph Based on a Cylindrical Focusing PVDF Antenna


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

We developed an optoacoustic tomograph with hand-held probe designed for optoacoustic imaging of biological tissues. The hand-held probe consists of a fiber-optic bundle for delivery of pulsed laser radiation to the studied object and a cylindrical focusing 64-element antenna for the detection of optoacoustic pulses. The capabilities of the tomograph to visualize the model blood vessels were studied experimentally using electronic and electronic–mechanical scanning. The achieved axial/lateral spatial resolution is 200/400 μm, the imaging depth is 18 mm, and the maximum B-scan acquisition rate is 10 Hz.

Sobre autores

P.V. Subochev

Institute of Applied Physics of the Russian Academy of Sciences

Autor responsável pela correspondência
Email: pavel.subochev@gmail.com
Rússia, Nizhny Novgorod

A. Postnikova

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Rússia, Nizhny Novgorod

A.V. Koval’chuk

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Rússia, Nizhny Novgorod

I. Turchin

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Rússia, Nizhny Novgorod

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer Science+Business Media, LLC, 2017