Biomedical Optoacoustic Tomograph Based on a Cylindrical Focusing PVDF Antenna


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

We developed an optoacoustic tomograph with hand-held probe designed for optoacoustic imaging of biological tissues. The hand-held probe consists of a fiber-optic bundle for delivery of pulsed laser radiation to the studied object and a cylindrical focusing 64-element antenna for the detection of optoacoustic pulses. The capabilities of the tomograph to visualize the model blood vessels were studied experimentally using electronic and electronic–mechanical scanning. The achieved axial/lateral spatial resolution is 200/400 μm, the imaging depth is 18 mm, and the maximum B-scan acquisition rate is 10 Hz.

Авторлар туралы

P.V. Subochev

Institute of Applied Physics of the Russian Academy of Sciences

Хат алмасуға жауапты Автор.
Email: pavel.subochev@gmail.com
Ресей, Nizhny Novgorod

A. Postnikova

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Ресей, Nizhny Novgorod

A.V. Koval’chuk

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Ресей, Nizhny Novgorod

I. Turchin

Institute of Applied Physics of the Russian Academy of Sciences

Email: pavel.subochev@gmail.com
Ресей, Nizhny Novgorod

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Springer Science+Business Media, LLC, 2017