A scintillation detector of pulsed soft X rays
- Autores: Nefedov Y.Y.1, Usenko P.L.1
-
Afiliações:
- Russian Federal Nuclear Center
- Edição: Volume 59, Nº 1 (2016)
- Páginas: 115-119
- Seção: General Experimental Technique
- URL: https://ogarev-online.ru/0020-4412/article/view/158916
- DOI: https://doi.org/10.1134/S0020441216010097
- ID: 158916
Citar
Resumo
A pulsed soft-X-ray detector consisting of a high-current photomultiplier tube and a thin-film polystyrene-based plastic scintillator was developed. Its characteristics were investigated by detecting quasimonochromatic 0.84-keV X-ray pulses. The dependence of the scintillator light output on the photon energy flux density was obtained. The average effective thickness of the surface non-scintillating (“dead”) layer affecting the spectral response of this detector type was estimated.