A Calorimeter Based on Peltier Modules
- Авторлар: Volkov A.S.1, Tyagunin A.V.1, Koposov G.D.1
-
Мекемелер:
- Northern (Arctic) Federal University named after M.V. Lomonosov
- Шығарылым: Том 61, № 5 (2018)
- Беттер: 766-768
- Бөлім: Laboratory Techniques
- URL: https://ogarev-online.ru/0020-4412/article/view/160435
- DOI: https://doi.org/10.1134/S0020441218040267
- ID: 160435
Дәйексөз келтіру
Аннотация
A two-chamber calorimeter with industrial TEC1-12706 Peltier modules used as heat-flow sensors and heating elements is described. The measured temperature dependence of the volt–watt sensitivity in the temperature range from –180 to 50°С is presented. The features of the measuring technique are described. The measured temperature dependence of the specific heat of ice is shown.
Авторлар туралы
A. Volkov
Northern (Arctic) Federal University named after M.V. Lomonosov
Хат алмасуға жауапты Автор.
Email: a.s.volkov@narfu.ru
Ресей, Arkhangel’sk, 163002
A. Tyagunin
Northern (Arctic) Federal University named after M.V. Lomonosov
Email: a.s.volkov@narfu.ru
Ресей, Arkhangel’sk, 163002
G. Koposov
Northern (Arctic) Federal University named after M.V. Lomonosov
Email: a.s.volkov@narfu.ru
Ресей, Arkhangel’sk, 163002
Қосымша файлдар