A highly sensitive gas microflowmeter
- 作者: Rumyantsev A.V.1, Kharjukov V.G.1
-
隶属关系:
- Immanuil Kant Baltic Federal University
- 期: 卷 60, 编号 2 (2017)
- 页面: 278-283
- 栏目: Laboratory Techniques
- URL: https://ogarev-online.ru/0020-4412/article/view/159675
- DOI: https://doi.org/10.1134/S0020441217020130
- ID: 159675
如何引用文章
详细
The design of a gas microflowmeter with the following characteristics is described: the gas flowrate measurement range is ~0–3.0 mg/s; the average sensitivity in this range is 19 V/(mg s–1); the range of recorded voltages is 25–85 V; and the error is ~2%. It is shown that when a convective heat exchange in calculations of the Reynolds and Nusselt numbers is described, the radius must serve as the characteristic size of a sphere but not its diameter.
作者简介
A. Rumyantsev
Immanuil Kant Baltic Federal University
编辑信件的主要联系方式.
Email: albert37@list.ru
俄罗斯联邦, Kaliningrad, 236041
V. Kharjukov
Immanuil Kant Baltic Federal University
Email: albert37@list.ru
俄罗斯联邦, Kaliningrad, 236041
补充文件