Calibration of cathode strip gains in multiwire drift chambers of the GlueX experiment
- Авторы: Berdnikov V.V.1, Somov S.V.1, Pentchev L.2, Somov A.2
-
Учреждения:
- National Research Nuclear University MEPhI (Moscow Engineering Physics Institute)
- Thomas Jefferson National Accelerator Facility
- Выпуск: Том 59, № 4 (2016)
- Страницы: 513-518
- Раздел: Nuclear Experimental Technique
- URL: https://ogarev-online.ru/0020-4412/article/view/159161
- DOI: https://doi.org/10.1134/S0020441216030027
- ID: 159161
Цитировать
Аннотация
A technique for calibrating cathode strip gains in multiwire drift chambers of the GlueX experiment is described. The accuracy of the technique is estimated based on Monte Carlo generated data with known gain coefficients in the strip signal channels. One of the four detector sections has been calibrated using cosmic rays. Results of drift chamber calibration on the accelerator beam upon inclusion in the GlueX experimental setup are presented.