A Method for Evaluating the Signal-to-Noise Ratio in Magnetic Resonance Images
- Авторы: Sergunova K.A.1, Akhmad E.S.1, Potrakhov N.N.2
-
Учреждения:
- Research and Practical Clinical Center of Diagnostics and Telemedicine Technologies, Department of Healthcare of Moscow
- Saint Petersburg Electrotechnical University “LETI”
- Выпуск: Том 53, № 3 (2019)
- Страницы: 207-210
- Раздел: Article
- URL: https://ogarev-online.ru/0006-3398/article/view/236156
- DOI: https://doi.org/10.1007/s10527-019-09910-3
- ID: 236156
Цитировать
Аннотация
Assessment of the signal-to-noise ratio as a control parameter for magnetic resonance imaging (MRI) systems is addressed. Experimental data were used to run a statistical analysis of noise components. Use of a multichannel radio-frequency receiver coil and determination of the coefficient of correction provided the basis of a method for assessing the signal-to-noise ratio of MRI scans.