Performance of a Right Ventricular Assist Rotary Pump in the Process of Biventricular Mechanical Circulatory Support
- 作者: Petukhov D.S.1, Telyshev D.V.1
-
隶属关系:
- National Research University of Electronic Technology
- 期: 卷 51, 编号 1 (2017)
- 页面: 33-36
- 栏目: Article
- URL: https://ogarev-online.ru/0006-3398/article/view/234872
- DOI: https://doi.org/10.1007/s10527-017-9679-6
- ID: 234872
如何引用文章
详细
Performance of a right ventricular assist rotary pump in the process of biventricular mechanical circulatory support was studied using a cardiovascular system model. Three modifications of the rotary pump and their pressure−flow curves were considered. It was shown that a normal pressure level and right ventricular unloading without collapse could be achieved using pumps with pressure−flow curves scaled to the pressure difference specific to pulmonary circulation.