Evaluation of Blood Plasma Coagulability by Laser Speckle Correlation
- 作者: Li L.1, Sytnik I.D.1, Gubarev F.A.1, Pekker Y.S.1,2
-
隶属关系:
- National Research Tomsk Polytechnic University
- Siberian State Medical University, Ministry of Health of the Russian Federation
- 期: 卷 52, 编号 3 (2018)
- 页面: 177-180
- 栏目: Article
- URL: https://ogarev-online.ru/0006-3398/article/view/235606
- DOI: https://doi.org/10.1007/s10527-018-9808-x
- ID: 235606
如何引用文章
详细
The process of blood plasma coagulation was studied using the laser speckle correlation technique. Coagulation times measured in samples of different volumes using the Minilab 701 commercial coagulometer and the laser speckle correlation technique were compared. It was shown that the laser speckle correlation technique provides comparable accuracy of measurement, while allowing the sample volume to be considerably reduced.