A Thermoelectric Device to Stop Bleeding
- Авторлар: Ismailov T.A.1, Yevdulov O.V.1, Ragimova T.A.1, Nabiev N.A.1
-
Мекемелер:
- Dagestan State Technical University
- Шығарылым: Том 53, № 2 (2019)
- Беттер: 92-96
- Бөлім: Article
- URL: https://ogarev-online.ru/0006-3398/article/view/236039
- DOI: https://doi.org/10.1007/s10527-019-09884-2
- ID: 236039
Дәйексөз келтіру
Аннотация
The design of a device for stopping bleeding is considered, in which the operative element is a thermoelectric battery (TEB) operating in cooling mode until the moment at which bleeding stops and in heating mode when the device is removed from the damage site. A model of the device based on solving the problem of thermal conductivity on phase transition is proposed. Numerical experimental results yielded a relationship between the time taken to form solid-phase blood 3 mm thick and the power and supply current of the TEB at different blood flow pressures in the wound zone. The device can use a standard TEB with power consumption of 40_80 W depending on the area of blood flow for procedures lasting 7-8 min.