Samples of commercial and noncommercial xenopericardial grafts of different origin and processing are analyzed. A full-scale test of uniaxial stretching of the material was used to evaluate the features of physical and mechanical properties of the samples, and a computer finite element method simulation was used to evaluate the degree of the object compression. The results were analyzed in terms of the capacity of selected pericardial grafts to serve as a valve apparatus of a transcatheter heart valve prosthesis.
Ключевые слова
Об авторах
E. Ovcharenko
Kemerovo Cardiology Center
Email: KlyshnikovK@gmail.com
Россия, Kemerovo
K. Klyshnikov
Kemerovo Cardiology Center
Автор, ответственный за переписку.
Email: KlyshnikovK@gmail.com
Россия, Kemerovo
T. Glushkova
Kemerovo Cardiology Center
Email: KlyshnikovK@gmail.com
Россия, Kemerovo
D. Nyshtaev
Tesis LLC
Email: KlyshnikovK@gmail.com
Россия, Moscow
Yu. Kudryavtseva
Kemerovo Cardiology Center
Email: KlyshnikovK@gmail.com
Россия, Kemerovo
G. Savrasov
Bauman Moscow State Technical University
Email: KlyshnikovK@gmail.com
Россия, Moscow
Дополнительные файлы