Tribological characteristics of hip joint endoprostheses
- 作者: Shuster L.S.1, Chertovskikh S.V.1, Minasov B.S.2, Yakupov R.R.2, Emaev I.I.1
-
隶属关系:
- Ufa State Aviation Technical University
- Bashkir State Medical University
- 期: 卷 37, 编号 1 (2016)
- 页面: 1-6
- 栏目: Article
- URL: https://ogarev-online.ru/1068-3666/article/view/227944
- DOI: https://doi.org/10.3103/S1068366616010165
- ID: 227944
如何引用文章
详细
We have studied the friction pairs for a head–insert kinematic node of human hip endoprostheses and rabbit hip endoprostheses, both normal (intact) and with an osteoarthritis model, in conditions of dry friction and lubrication with biological medium. We have discovered that the rabbit hip with an osteoarthritis model exhibits deterioration of tribological characteristics. From the tribological point of view, the most suitable friction pairs for human hip endoprostheses are zirconium ceramics–aluminum ceramics, Oxinium–aluminum ceramics, aluminum ceramics–aluminum ceramics, and aluminum ceramics–polyethylene, which have shown the smallest coefficient of friction, the lowest adhesion bond strength when sheared, and minimal dependence of these parameters on the value of external load.
作者简介
L. Shuster
Ufa State Aviation Technical University
Email: chertovskikh@mail.ru
俄罗斯联邦, ul. K. Marksa 12, Ufa, Bashkortostan, 450000
S. Chertovskikh
Ufa State Aviation Technical University
编辑信件的主要联系方式.
Email: chertovskikh@mail.ru
俄罗斯联邦, ul. K. Marksa 12, Ufa, Bashkortostan, 450000
B. Minasov
Bashkir State Medical University
Email: chertovskikh@mail.ru
俄罗斯联邦, ul. Lenina 3, Ufa, Bashkortostan, 450000
R. Yakupov
Bashkir State Medical University
Email: chertovskikh@mail.ru
俄罗斯联邦, ul. Lenina 3, Ufa, Bashkortostan, 450000
I. Emaev
Ufa State Aviation Technical University
Email: chertovskikh@mail.ru
俄罗斯联邦, ul. K. Marksa 12, Ufa, Bashkortostan, 450000
补充文件