Reduction of Effect of Concomitant Gradients in Low Magnetic Field MRI via Optimization of Gradient Magnetic System


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

The effect of concomitant magnetic fields emerging in conjunction with encoding gradients, which is important in the process of the magnetic resonance imaging in low fields, has been considered. The manifestations of concomitant magnetic fields in a concrete gradient system, namely in the system of two coaxial gradient coils, have been thoroughly analyzed. It has been suggested to improve the gradient system via optimization of the interspace between coils on the basis of the standard criterion of the minimum of root-mean-square deviation of the encoding field dependence from a linear one. It has been shown that the optimal interspace is not the Maxwell condition.

Sobre autores

Vladimir Chizhik

Faculty of Physics, Saint Petersburg State University

Email: vfrolovv@bk.ru
Rússia, Saint-Peterburg

Vyacheslav Frolov

Faculty of Physics, Saint Petersburg State University

Autor responsável pela correspondência
Email: vfrolovv@bk.ru
Rússia, Saint-Peterburg

Pavel Kupryanov

Faculty of Physics, Saint Petersburg State University

Email: vfrolovv@bk.ru
Rússia, Saint-Peterburg

Konstantin Tyutyukin

Faculty of Physics, Saint Petersburg State University

Email: vfrolovv@bk.ru
Rússia, Saint-Peterburg

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer-Verlag GmbH Austria, 2017