Time Interval Measurement in the Picosecond Accuracy


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

A model and a laboratory method were developed for determining the major systematic errors affecting time interval measurement (TIM) accuracy. The laboratory method investigates the optimum trigger level for the high slew rate timing pulses that achieves the highest TIM accuracy. The overall errors due to the measurement setup and the trigger level, at which the measurements are executed, are the most considerable sources of systematic errors. Previously these error sources were modeled. In this study the model was modified for estimating more precisely the systematic errors contributed by the measurement setup. Also a laboratory method was experienced to determine the optimum trigger level for the timing pulses. Applying these methods achieve a TIM accuracy in the picosecond range.

作者简介

Gihan Hamza

National Institute for Standards

编辑信件的主要联系方式.
Email: gihan_gomah@yahoo.com
埃及, Giza

补充文件

附件文件
动作
1. JATS XML