A Controlled Magnetic Clamp
- Autores: Edelman V.S.1
-
Afiliações:
- Kapitsa Institute of Physical Problems
- Edição: Volume 61, Nº 5 (2018)
- Páginas: 751-752
- Seção: Laboratory Techniques
- URL: https://ogarev-online.ru/0020-4412/article/view/160424
- DOI: https://doi.org/10.1134/S0020441218050172
- ID: 160424
Citar
Resumo
A clamp for establishing a thermal contact between cryostat components in a vacuum is described. The contact is established by pressing heat conductors to each other using a permanent magnet and is broken or established by applying a current pulse of the required polarity to the winding around an iron magnetic core. The device is held in the open state by a second permanent magnet. Through the pressed contacts, a copper block with a mass of ∼600 g is cooled in the vacuum from the liquid-nitrogen temperature to the liquid- helium temperature within 4–5 h.