A Prototype of a Ring Neutron Detector
- Authors: Baskov V.A.1, Govorkov B.B.1, Poliansky V.V.1
-
Affiliations:
- Lebedev Physical Institute, Russian Academy of Sciences
- Issue: Vol 62, No 3 (2019)
- Pages: 305-308
- Section: Nuclear Experimental Technique
- URL: https://ogarev-online.ru/0020-4412/article/view/160691
- DOI: https://doi.org/10.1134/S0020441219030035
- ID: 160691
Cite item
Abstract
When the half-ring of a prototype neutron detector with an inner radius of 50 mm is calibrated using a 60Co ionizing-radiation source, its relative detection efficiency gradually decreases from ~15% at the center of the half-ring to ~5% at its ends. When calibrated using cosmic muons, the relative detection efficiency is approximately 100% over the entire detector volume. The presence of zones with reduced and increased efficiency values are associated with the intricate detector geometry.
About the authors
V. A. Baskov
Lebedev Physical Institute, Russian Academy of Sciences
Author for correspondence.
Email: baskov@x4u.lebedev.ru
Russian Federation, Moscow, 119991
B. B. Govorkov
Lebedev Physical Institute, Russian Academy of Sciences
Email: baskov@x4u.lebedev.ru
Russian Federation, Moscow, 119991
V. V. Poliansky
Lebedev Physical Institute, Russian Academy of Sciences
Email: baskov@x4u.lebedev.ru
Russian Federation, Moscow, 119991
Supplementary files