A Device for Precise Isothermal Sample Annealing in an Accelerator Chamber


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

A device has been developed for isothermal sample annealing in the chamber of an accelerator facility. The application of the nuclear reaction analysis is allowed during the sample annealing. Flowing liquid nitrogen and a resistive heater are used for the sample cooling and heating. Using the device, sample annealing is carried out in the temperature range of 110–500 K and the temperature is maintained with an accuracy of 1 K. The obtained characteristics meet the requirements of precision experiments on deuterium diffusion in solids, including at cryogenic temperatures.

About the authors

V. N. Rybakov

Mikheev Institute of Metal Physics, Ural Branch, Russian Academy of Sciences

Email: obuhovsvii@imp.uran.ru
Russian Federation, Yekaterinburg, 620108

A. V. Suchkov

Mikheev Institute of Metal Physics, Ural Branch, Russian Academy of Sciences

Email: obuhovsvii@imp.uran.ru
Russian Federation, Yekaterinburg, 620108

V. B. Vykhodets

Mikheev Institute of Metal Physics, Ural Branch, Russian Academy of Sciences

Email: obuhovsvii@imp.uran.ru
Russian Federation, Yekaterinburg, 620108

S. I. Obuhov

Mikheev Institute of Metal Physics, Ural Branch, Russian Academy of Sciences

Author for correspondence.
Email: obuhovsvii@imp.uran.ru
Russian Federation, Yekaterinburg, 620108

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2019 Pleiades Publishing, Inc.