Scientific Grounds for the Design of Electrical Impedance Systems for Monitoring the Parameters of Central Hemodynamics and Respiration


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

Among the instrumented methods for assessing the volume characteristics of cardiac activity, ultrasound, MRI, and CT with contrast are well known. Conventional methods do not provide for measurement of the parameters of central hemodynamics and respiration in monitoring conditions. The possibility of multichannel electrical impedance technology allows the stroke volume and fractional output of the heart to be determined and the displacement of the ventricular walls to be visualized in real time with an accuracy no worse than 1-2 mm. This article considers questions of the implementation of techniques for computing stroke and minute volumes of the circulation and displacements of the heart chamber by solving the electrical impedance measurement problem. Results of mathematical and physical modeling of cardiac activity based on precordial measurements are presented.

Sobre autores

I. Sergeev

Russian Scientific and Research Institute for Medical Engineering, Federal Service for Surveillance in Healthcare; Bauman Moscow State Technical University

Autor responsável pela correspondência
Email: sergeevik@rambler.ru
Rússia, Moscow; Moscow

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer Science+Business Media, LLC, part of Springer Nature, 2019